Temperature Simulator & Measurement SIKA MC 50.2 & MC 75.2
Spesifikasi Temperature Simulator & Measurement SIKA MC 50.2 & MC 75.2
Alat Ukur Kalibrasi
Type | MC 50.2 | MC 75.2 |
Resistance thermometer (RTD) | ||
RTD signal generation and measurement | Pt50, Pt100, Pt200, Pt500, Pt1000, Cu10, Cu50, Ni100, Ni120, Ni1000 | |
Ω generator function | 0...4000 Ω | |
Ω measurement function | 0...4000 Ω | |
Accuracy (of rdg. + const.) | ±0.012 % | |
Selectable temperature unit | °C / °F | |
Measurement of multi-wire | 2 / 3 / 4 | |
Thermocouples (TC) | ||
TC signal generation and measurement | J, K, T, R, S, B, N, E, U, L | |
mV generator function | 0...100 mV | |
mV measurement function | 0...100 mV | |
Accuracy (of rdg. + const.) | ±0.013 % | |
Selectable temperature unit | °C / °F | |
Internal comparison point | ±0.3 °C | |
Current (mA) | ||
Loop current signal generation | 0(4)...24 mA | |
Loop current signal measurement | 0(4)...50 mA | |
Accuracy (of rdg. + const.) | ±0.0175 % | |
Current loop supply | 24 V ±10 %, 22 mA | |
HART compatible internal loop resistor | 250 Ω | |
Voltage (V) | ||
Voltage signal generation | 0...20 V | |
Voltage signal measurement | 0...50 V | |
Accuracy (of rdg. + const.) | ±0.015 % | |
Continuity (O / C) | ||
Continuity measurement | O / C | |
Switching threshold "open" | 1 kΩ | |
Frequency and pulse (Hz) | ||
Frequency and pulse signal generation | 0.01 Hz...10 kHz | |
Frequency and pulse signal measurement | 0.01 Hz...20 kHz | |
Accuracy (of rdg. + const.) | ±0.005 % | |
Pressure signals (bar) | ||
Pressure measurement with external pressure module | √ | |
Editable pressure units | √ | |
Connection via DIN socket | 5-pin |
MC 50.2
- Operator guidance
Separate channels for parallel signal processing
Menu with pull-down windows
Programming and control via PC Graphic display of connection options
Configuration files for test adjustments (10)
- Battery supply / Operating time Approx. 25 h with Auto-Power off
MC 75.2
- Operator guidance
Separate channels for parallel signal processing Menu with pull-down windows
Programming and control via PC Graphic display of connection options
Configuration files for test adjustments (10)
Test report generation / Calibration routines (10)
- Battery supply / Operating time Approx. 20 h with Auto-Power off
- Test certificate
- Software (optional)
►Pressure measurement
MC 75.2 with external pressure modules
For universal on-site use, the MC 75 2 simulator is capable of measuring different pressures from -1 bar to 1000 bar.
The EPM plug-in modules are based on intelligent sensor technology with storage for measuring range and accuracy. Connection takes place via a DIN socket with Plug & Play function.
For the best measuring results with high accuracy, various absolute or relative pressure ranges in three precision classes are available.
EXTERNAL PRESSURE MODULE :
Technical data | Type EPM |
Measuring rate | 400 measurements/sec. |
Digital signals | RS 485 |
Electrical connection | 5-pin 1 m shielded connecting cable |
Medium temperature | 10…40 °C |
Pressure connection | G¼ stainless steel 1.4404 |
Degree of protection | IP65 |
Dimensions | Approx. D = 30 mm, L = 110 mm |
Weight | Approx. 140 g |
► EPM VERSION AVAILABLE:
Type EPM | Model A | Model B | Model C | |
Precision | ±0.05 % full scale | ±0.025 % full scale | ±0.01 % full scale | Max. pressure range |
1 | -1...1 bar rel. (Pr) | 2 bar | ||
0...1 bar abs. (Paa) | ||||
3 | -1...3 bar rel. (Pr) | 5 bar | ||
0…3 bar abs. (Paa) | ||||
4 | -1…4 bar rel. (PR) | -1…4 bar rel. (PA) | 20 bar | |
0…4 bar abs. (PAA) | 0…4 bar abs. (PAA) | |||
7 | -1…7 bar rel. (PR) | -1…7 bar rel. (PA) | 20 bar | |
0…7 bar abs. (PAA) | 0…7 bar abs. (PAA) | |||
10 | -1…10 bar rel. (Pr) | -1…10 bar rel. (PA) | -1…10 bar rel. (Pa) | 20 bar |
0…10 bar abs. (Paa) | 0…10 bar abs. (PAA) | 0…10 bar abs. (Paa) | ||
12 | -1…12 bar rel. (PR) | -1…12 bar rel. (PA) | 60 bar | |
0…12 bar abs. (PAA) | 0…12 bar abs. (PAA) | |||
20 | -1…20 bar rel. (PR) | -1…20 bar rel. (PA) | 60 bar | |
0…20 bar abs. (PAA) | 0…20 bar abs. (PAA) | |||
30 | -1…30 bar rel. (Pr) | -1…30 bar rel. (PA) | -1…30 bar rel. (Pa) | 60 bar |
0…30 bar abs. (Paa) | 0…30 bar abs. (PAA) | 0…30 bar abs. (Paa) | ||
40 | -1…40 bar rel. (PA) | -1…40 bar rel. (PA) | 200 bar | |
0…40 bar abs. (PAA) | 0…40 bar abs. (PAA) | |||
70 | 0…70 bar rel. (PA) | 0…70 bar rel. (PA) | 200 bar | |
0…70 bar abs. (PAA) | 0…70 bar abs. (PAA) | |||
100 | 0…100 bar rel. (Pa) | 0…100 bar rel. (PA) | 0…100 bar rel. (Pa) | 200 bar |
0…100 bar abs. (Paa) | 0…100 bar abs. (PAA) | 0…100 bar abs. (Paa) | ||
120 | 0…120 bar rel. (PA) | 0…120 bar rel. (PA) | 400 bar | |
0…120 bar abs. (PAA) | 0…120 bar abs. (PAA) | |||
135 | 0…135 bar rel. (PA) | 0…135 bar rel. (PA) | 400 bar | |
0…135 bar abs. (PAA) | 0…135 bar abs. (PAA) | |||
160 | 0…160 bar rel. (PA) | 0…160 bar rel. (PA) | 400 bar | |
0…160 bar abs. (PAA) | 0…160 bar abs. (PAA) | |||
200 | 0…200 bar rel. (PA) | 0…200 bar rel. (PA) | 400 bar | |
0…200 bar abs. (PAA) | 0…200 bar abs. (PAA) | |||
300 | 0…300 bar rel. (Pa) | 0…300 bar rel. (PA) | 0…300 bar rel. (Pa) | 400 bar |
0…300 bar abs. (Paa) | 0…300 bar abs. (PAA) | 0…300 bar abs. (Paa) | ||
400 | 0…400 bar rel. (PA) | 0…400 bar rel. (PA) | 1000 bar | |
0…400 bar abs. (PAA) | 0…400 bar abs. (PAA) | |||
700 | 0…700 bar rel. (Pa) | 0…700 bar rel. (PA) | 0…700 bar rel. (Pa) | 1000 bar |
0…700 bar abs. (Paa) | 0…700 bar abs. (PAA) | 0…700 bar abs. (Paa) | ||
1000 | 0…1000 bar rel. (Pa) | 0…1000 bar rel. (PA) | 0…1000 bar rel. (Pa) | 1000 bar |
0…1000 bar abs. (Paa) | 0…1000 bar abs. (PAA) | 0…1000 bar abs. (Paa) |
PR: Relative pressure measuring cell, ambient pressure as zero point
PAA: Absolute pressure measuring cell, vacuum as zero point
PA: Absolute pressure measuring cell, ambient pressure as zero point