Alat Ukur Kalibrasi
Type
|
MC 50.2
|
MC 75.2
|
Resistance thermometer (RTD)
|
RTD signal
generation and measurement
|
Pt50, Pt100, Pt200, Pt500, Pt1000, Cu10, Cu50, Ni100, Ni120, Ni1000
|
Ω generator function
|
0...4000 Ω
|
Ω measurement function
|
0...4000 Ω
|
Accuracy (of rdg. + const.)
|
±0.012 %
|
Selectable temperature unit
|
°C / °F
|
Measurement of multi-wire
|
2 / 3 / 4
|
Thermocouples (TC)
|
TC signal generation and measurement
|
J, K, T, R, S, B, N, E, U, L
|
mV generator function
|
0...100 mV
|
mV measurement function
|
0...100 mV
|
Accuracy (of rdg. + const.)
|
±0.013 %
|
Selectable temperature unit
|
°C / °F
|
Internal comparison point
|
±0.3 °C
|
Current (mA)
|
Loop current signal generation
|
0(4)...24 mA
|
Loop current signal measurement
|
0(4)...50 mA
|
Accuracy (of rdg. + const.)
|
±0.0175 %
|
Current loop supply
|
24 V ±10 %, 22 mA
|
HART compatible internal loop resistor
|
250 Ω
|
Voltage (V)
|
Voltage signal generation
|
0...20 V
|
Voltage signal measurement
|
0...50 V
|
Accuracy (of rdg. + const.)
|
±0.015 %
|
Continuity (O / C)
|
Continuity measurement
|
O / C
|
Switching threshold "open"
|
1 kΩ
|
Frequency and pulse (Hz)
|
Frequency and pulse signal generation
|
0.01 Hz...10 kHz
|
Frequency and pulse signal measurement
|
0.01 Hz...20 kHz
|
Accuracy (of rdg. + const.)
|
±0.005 %
|
Pressure signals (bar)
|
Pressure measurement
with external pressure module
|
|
√
|
Editable pressure units
|
|
√
|
Connection via DIN socket
|
|
5-pin
|
MC 50.2
Separate channels for parallel signal processing
Menu with pull-down windows
Programming and control via PC Graphic display of connection options
Configuration files for test adjustments (10)
- Battery supply / Operating time Approx. 25 h with Auto-Power off
MC 75.2
Separate channels for parallel signal processing Menu with pull-down windows
Programming and control via PC Graphic display of connection options
Configuration files for test adjustments (10)
Test report generation / Calibration routines (10)
- Battery supply / Operating time Approx. 20 h with Auto-Power off
- Test certificate
- Software (optional)
►Pressure measurement
MC 75.2 with external pressure modules
For universal on-site use, the MC 75 2 simulator is capable of measuring different pressures from -1 bar to 1000 bar.
The EPM plug-in modules are based on intelligent sensor technology with storage for measuring range and accuracy. Connection takes place via a DIN socket with Plug & Play function.
For the best measuring results with high accuracy, various absolute or relative pressure ranges in three precision classes are available.
EXTERNAL PRESSURE MODULE :
Technical data
|
Type EPM
|
Measuring rate
|
400 measurements/sec.
|
Digital signals
|
RS 485
|
Electrical connection
|
5-pin 1 m shielded connecting cable
|
Medium temperature
|
10…40 °C
|
Pressure connection
|
G¼ stainless steel 1.4404
|
Degree of protection
|
IP65
|
Dimensions
|
Approx. D = 30 mm, L = 110 mm
|
Weight
|
Approx. 140 g
|
► EPM VERSION AVAILABLE:
Type EPM
|
Model A
|
Model B
|
Model C
|
|
Precision
|
±0.05 % full scale
|
±0.025 % full scale
|
±0.01 % full scale
|
Max. pressure range
|
1
|
-1...1 bar rel. (Pr)
|
|
|
2 bar
|
0...1 bar abs. (Paa)
|
|
|
|
3
|
-1...3 bar rel. (Pr)
|
|
|
5 bar
|
0…3 bar abs. (Paa)
|
|
|
|
4
|
-1…4 bar rel. (PR)
|
-1…4 bar rel. (PA)
|
|
20 bar
|
0…4 bar abs. (PAA)
|
0…4 bar abs. (PAA)
|
|
|
7
|
-1…7 bar rel. (PR)
|
-1…7 bar rel. (PA)
|
|
20 bar
|
0…7 bar abs. (PAA)
|
0…7 bar abs. (PAA)
|
|
|
10
|
-1…10 bar rel. (Pr)
|
-1…10 bar rel. (PA)
|
-1…10 bar rel. (Pa)
|
20 bar
|
0…10 bar abs. (Paa)
|
0…10 bar abs. (PAA)
|
0…10 bar abs. (Paa)
|
|
12
|
-1…12 bar rel. (PR)
|
-1…12 bar rel. (PA)
|
|
60 bar
|
0…12 bar abs. (PAA)
|
0…12 bar abs. (PAA)
|
|
|
20
|
-1…20 bar rel. (PR)
|
-1…20 bar rel. (PA)
|
|
60 bar
|
0…20 bar abs. (PAA)
|
0…20 bar abs. (PAA)
|
|
|
30
|
-1…30 bar rel. (Pr)
|
-1…30 bar rel. (PA)
|
-1…30 bar rel. (Pa)
|
60 bar
|
0…30 bar abs. (Paa)
|
0…30 bar abs. (PAA)
|
0…30 bar abs. (Paa)
|
|
40
|
-1…40 bar rel. (PA)
|
-1…40 bar rel. (PA)
|
|
200 bar
|
0…40 bar abs. (PAA)
|
0…40 bar abs. (PAA)
|
|
|
70
|
0…70 bar rel. (PA)
|
0…70 bar rel. (PA)
|
|
200 bar
|
0…70 bar abs. (PAA)
|
0…70 bar abs. (PAA)
|
|
|
100
|
0…100 bar rel. (Pa)
|
0…100 bar rel. (PA)
|
0…100 bar rel. (Pa)
|
200 bar
|
0…100 bar abs. (Paa)
|
0…100 bar abs. (PAA)
|
0…100 bar abs. (Paa)
|
|
120
|
0…120 bar rel. (PA)
|
0…120 bar rel. (PA)
|
|
400 bar
|
0…120 bar abs. (PAA)
|
0…120 bar abs. (PAA)
|
|
|
135
|
0…135 bar rel. (PA)
|
0…135 bar rel. (PA)
|
|
400 bar
|
0…135 bar abs. (PAA)
|
0…135 bar abs. (PAA)
|
|
|
160
|
0…160 bar rel. (PA)
|
0…160 bar rel. (PA)
|
|
400 bar
|
0…160 bar abs. (PAA)
|
0…160 bar abs. (PAA)
|
|
|
200
|
0…200 bar rel. (PA)
|
0…200 bar rel. (PA)
|
|
400 bar
|
0…200 bar abs. (PAA)
|
0…200 bar abs. (PAA)
|
|
|
300
|
0…300 bar rel. (Pa)
|
0…300 bar rel. (PA)
|
0…300 bar rel. (Pa)
|
400 bar
|
0…300 bar abs. (Paa)
|
0…300 bar abs. (PAA)
|
0…300 bar abs. (Paa)
|
|
400
|
0…400 bar rel. (PA)
|
0…400 bar rel. (PA)
|
|
1000 bar
|
0…400 bar abs. (PAA)
|
0…400 bar abs. (PAA)
|
|
|
700
|
0…700 bar rel. (Pa)
|
0…700 bar rel. (PA)
|
0…700 bar rel. (Pa)
|
1000 bar
|
0…700 bar abs. (Paa)
|
0…700 bar abs. (PAA)
|
0…700 bar abs. (Paa)
|
|
1000
|
0…1000 bar rel. (Pa)
|
0…1000 bar rel. (PA)
|
0…1000 bar rel. (Pa)
|
1000 bar
|
0…1000 bar abs. (Paa)
|
0…1000 bar abs. (PAA)
|
0…1000 bar abs. (Paa)
|
|
PR: Relative pressure measuring cell, ambient pressure as zero point
PAA: Absolute pressure measuring cell, vacuum as zero point
PA: Absolute pressure measuring cell, ambient pressure as zero point