© - Powered by Indotrading.
Type
TP 37200E.2
TP 37165E.2
Control sensor
Switchable internal / external
Hysteresis
±0.25 °C / ±0.025 °C (internal / external)
Temperature range*
-55...200 °C
-35...165 °C
Dry block
Accuracy Stability
±0.2 °C
±0.010 °C / ±0.005 °C (internal / external)
Measurement zone
110...150 mm
Technical data
Optional for temperature calibrators
TP 37200E.2 / TP 37165E.2*
Device under test input - Resistance thermometers
Number of channels
2
Connection
4x 4 mm safety socket per channel
Type of connection
2-, 3-, 4-wire technology
Resistance range
0…400 Ω
Accuracy Pt100
±0.05 °C
Accuracy Pt1000
±0.09 °C
Device under test input - Thermocouple
2x thermocouple socket (Mini)
Measuring range
0…100 mV
Accuracy cold junction
±0.3 °C
Accuracy
±0.08 °C
→ Type K
→ Type J
±0.07 °C
→ Type N
±0.13 °C
→ Type E
±0.06 °C
→ Type T
→ Type R
±0.78 °C
→ Type S
±0.73 °C
Standard signal input (Current)
1
4 mm safety socket
0…24 mA
0.01 % full scale
Standard signal input (Voltage)
0…12 VDC
Switch test
Transmitter supply
Output current
Max. 24 mA
Output voltage
24 VDC
Alat Ukur Kalibrasi
Untuk keterangan lebih lanjut, silahkan download PDF ini :
Rp. CALL